K71000 Automatic PMCC Flash Point Analyser

The K71000 Automatic PMCC Flash Point Analyser is a perfect union of next-generation technology with traditional robust quality for determination of flash point of biodiesel, distillate fuels, new lubricating oils, residual fuel oils, cutback residual, used lubricating oils, mixtures of petroleum liquids with solids, and petroleum liquids that tend to form a surface film during testing.

K71000 Automatic PMCC Flash Point Analyser
Request Info/Quote

K71000 - Automatic Pensky-Martens Closed Cup Flash Point Tester

Conforms to the specifications of

ASTM D93 Procedure A, B and C; IP 34; ISO 2719; DIN EN 22719; JIS K2265; NF M 07-019

Key Features

  • Flash point detection by thermocouple and ionization ring
  • Electric or Gas ignition: Software selectable,user friendly Manual Switching
  • Flash point operation range between ambient and 405°C
  • A three (3) position mechanical lift system for the cover and motor assembly is fully automated and software selectable: Open - Clean -Test for one touch positioning of the test cup
  • Integrated Dual Fan System directly cools test cup and surrounding environment
  • The number of user programs including a quick safely test from ambient, puts the flash point result into the EFP of the official run, and prompts user to refresh the sample, virtually assuring no fires ever occur.
  • 8.4” LCD Touch Screen Interface allows for easy viewing and navigation
  • Over 65,000 results of data can be stored
  • Automatic Barometric Pressure Correction
  • Fire Suppression System floods instrument with inert gas in the event of a fire

Specifications

Heating Rate:In accordance with ASTM D93 procedure A, B and C
Stirring Rate:0 to 300 ± 5 RPM
Cycle Time:5 minutes as per rapid fan cooling system Ignition Test Frequency: User selectable on per method basis
Temperature Calibration:Three Types as follows:
2 Point Sample Temperature
Multipoint (>2) Temperature Calibration
CRM Calibration – Built-In Software function to automatically correct as per CRM tested

Communication:LIMS Connectivity via Ethernet and RS 232 ports. Two (2) USB ports (1–Front , 2–Back)
Security:Three (3) level password protection
Request Info/Quote